Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Novel Potentiometric Sensor for Monitoring Beryllium Based on Naphto-9-crown-3
Mohammad Reza GANJALIAzadeh DAFTARIMajid FAAL-RASTEGARAbolghasem MOGHIMI
Author information
JOURNAL FREE ACCESS

2003 Volume 19 Issue 3 Pages 353-356

Details
Abstract

A novel poly(vinyl chloride) (PVC) membrane electrode based on naphto-9-crown-3 was prepared and tested for the selective detection of beryllium ions. A suitable lipophilicity of the carrier and appropriate coordination ability were found to be essential for designing an electrode with good response characteristics. A PVC membrane with 9% naphtho-9-crown-3 carrier, 58% o-NPOE plasticizer, 3% tetraphenylborate anionic excluder and 30% poly(vinyl chloride) satisfied these requirements. The proposed sensor displayed a linear response to beryllium over a wide concentration range of 1.0 × 10-1-8.0 × 10-6 M with a Nernstian slope of 29.5 mV per decade. The electrode showed very short response time (<15 s) and could be used in the pH range 3.5 - 9.0. The selectivity coefficient for alkali, alkaline earth, transition and heavy metal ions was smaller than 4.0 × 10-4. The sensor was successfully used as an indicator electrode in the potentiometric titration of Be2+ with EDTA. The proposed Be(II) sensor was also used for the determination of Be2+ ions in binary mixtures.

Content from these authors
© 2003 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top