Magnetic Resonance in Medical Sciences
Online ISSN : 1880-2206
Print ISSN : 1347-3182
ISSN-L : 1347-3182
Major Papers
3D MR Cisternography to Identify Distal Dural Rings: Comparison of 3D-CISS and 3D-SPACE Sequences
Yoshiyuki WATANABEAkari MAKIDONOMiho NAKAMURAYukihisa SAIDA
Author information
JOURNAL OPEN ACCESS

2011 Volume 10 Issue 1 Pages 29-32

Details
Abstract

Purpose: The distal dural ring (DDR) is an anatomical landmark used to distinguish intra- and extradural aneurysms. We investigated identification of the DDR using 2 three-dimensional (3D) magnetic resonance (MR) cisternography sequences-3D constructive interference in steady state (CISS) and 3D sampling perfection with application optimized contrasts using different flip angle evolutions (SPACE)-at 3.0 tesla.
Methods: Ten healthy adult volunteers underwent imaging with 3D-CISS, 3D-SPACE, and time-of-flight (TOF) MR angiography (TOF-MRA) sequences at 3.0T. We analyzed DDR identification and internal carotid artery (ICA) signal intensity and classified the shape of the carotid cave.
Results: We identified the DDR using both 3D-SPACE and 3D-CISS, with no significant difference between the sequences. Visualization of the outline of the ICA in the cavernous sinus (CS) was significantly clearer with 3D-SPACE than 3D-CISS. In the CS and petrous portions, signal intensity was lower with 3D-SPACE, and the flow void was poor with 3D-CISS in some subjects.
Conclusion: We identified the DDR with both 3D-SPACE and 3D-CISS, but the superior contrast of the ICA in the CS using 3D-SPACE suggests the superiority of this sequence for evaluating the DDR.

Content from these authors
© 2011 by Japanese Society for Magnetic Resonance in Medicine
Previous article Next article
feedback
Top