Magnetic Resonance in Medical Sciences
Online ISSN : 1880-2206
Print ISSN : 1347-3182
ISSN-L : 1347-3182
Major Papers
Standardizing Display Conditions of Diffusion-weighted Images Using Concurrent b0 Images: A Multi-vendor Multi-institutional Study
Makoto SASAKIMasahiro IDAKei YAMADAYoshiyuki WATANABEMieko MATSUI
Author information
JOURNAL OPEN ACCESS

2007 Volume 6 Issue 3 Pages 133-137

Details
Abstract

Purpose: To establish a practical method that uses concurrent b0 images to standardize the display conditions for diffusion-weighted images (DWI) that vary among institutions and interpreters.
Method: Using identical parameters, we obtained DWI for 12 healthy volunteers at 4 institutions using 4 MRI scanners from 3 vendors. Three operators manually set the window width for the images equal to the signal intensity of the normal-appearing thalamus on b0 images and set the window level at half and then exported the images to 8-bit gray-scale images. We calculated the mean pixel values of the brain objects in the images and examined the variation among scanners, operators, and subjects.
Result: Following our method, the DWI of the 12 subjects obtained using the 4 different scanners had nearly identical contrast and brightness. The mean pixel values of the brain on the exported images among the operators and subjects were not significantly different, but we found a slight, significant difference among the scanners.
Conclusion: Determining DWI display conditions by using b0 images is a simple and practical method to standardize window width and level for evaluating diffusion abnormalities and decreasing variation among institutions and operators.

Content from these authors
© 2007 by Japanese Society for Magnetic Resonance in Medicine
Previous article Next article
feedback
Top